The Health Products and Food Branch (HPFB) posts on the Health Canada web site safety alerts, public health advisories, press releases and other notices as a service to health professionals, consumers, and other interested parties. These advisories may be prepared with Directorates in the HPFB which includes pre-market and post-market areas as well as market authorization holders and other stakeholders. Although the HPFB grants market authorizations or licenses for therapeutic products, we do not endorse either the product or the company. Any questions regarding product information should be discussed with your health professional.

This is duplicated text of a letter from Baxter Corporation.
Contact the company for a copy of any references, attachments or enclosures.


Health Canada Endorsed Important Safety Information on the Colleague Volumetric Infusion Pump

15 December 2005

URGENT DEVICE CORRECTION

Re: COLLEAGUE VOLUMETRIC INFUSION PUMP, PRODUCT CODES: 2M8151 & 2M8151R;
COLLEAGUE 3 VOLUMETRIC INFUSION PUMP, PRODUCT CODES: 2M8153 & 2M8153R;
COLLEAGUE CX VOLUMETRIC INFUSION PUMP, PRODUCT CODES: 2M8161 & 2M8161R;
COLLEAGUE 3 CX VOLUMETRIC INFUSION PUMP, PRODUCT CODES: 2M8163 & 2M8163R
FRENCH COLLEAGUE VOLUMETRIC PUMP, PRODUCT CODES: DNM8151 & DNM8151R
FRENCH COLLEAGUE 3 VOLUMETRIC PUMP, PRODUCT CODES: DNM8153, & DNM8153R

Dear Director of Nursing:

Baxter Corporation is sending this communication to notify you of an Urgent Device Correction related to the COLLEAGUE Volumetric Infusion Pump. During the investigation into resolution of issues previously communicated to customers, Baxter has identified additional issues that we want to bring to your attention. These issues fall into the following categories:

  • Battery Undercharging
  • Generation of False Air Detected Alarm due to IV Administration Set Tugging
  • Gearbox Wear
  • Underinfusion
  • Non-Detection of Upstream Occlusion

Additional information to help users avoid conditions that may interrupt therapy, along with other guidance on the usage of your COLLEAGUE pump, is described in the "Battery Usage Guide" (Attachment 1) and "Infusion Management Guide" (Attachment 2). These attachments should be provided to all users of the COLLEAGUE pump.

Battery Undercharging:

Baxter has received one report of a patient death that may be associated with an undercharged battery.

Users of COLLEAGUE pumps should be aware that if the pump's batteries are not charged continuously for 12 hours after a Battery Low alert or Battery Depleted alarm occurs, the remaining operating time after a subsequent Battery Low alert can be less than 30 minutes. This will be followed by a Battery Depleted alarm with both an audible and visual notification and the infusion(s) will stop. In order to restart the infusion(s), users must plug the pump into a power outlet immediately.

In order to avoid the possibility of interruption or cessation of therapy, loss of configuration memory, and/or device failure, it is important that you follow the charging instructions listed in the Operator's Manual prior to using the pump under battery operation. Please also refer to the "Battery Usage Guide" (Attachment 1) for additional information on proper battery usage and maintenance.

Generation of False Air Detected Alarm due to IV Administration Set Tugging:

Baxter has received one report of a patient death that may be associated with an interruption of therapy due to a false Air Detected alarm.

Pulling or tugging on the IV administration set, between the pump channel and the patient, may cause a false Air Detected alarm, which will cause the pump to stop infusing and issue an audible and visual notification. In order to reduce the potential for this situation to occur, first select an appropriate length administration set. Before loading the set into the pump, position the keyed slide clamp at an appropriate location along the administration set to ensure that there is adequate length of tubing between the patient and the pump to reduce tugging on the set during activities such as moving the patient from one bed to another, or transportation of the patient from one facility location to another.

Gearbox Wear:

Baxter has identified that worn gearbox components can result in Failure Codes 812:02 and 812:05, which will result in an audible and visual alarm and the interruption of therapy. As this is associated with age and wear on the pump, older pumps are more prone to exhibit these failures. As stated in the Operator and Service Manuals, if you experience these failure codes immediately remove the pump from service and have the pump evaluated by Baxter-trained service personnel. Baxter is in the process of developing actions to address this issue and will notify you once they are finalized.

Underinfusion:

Baxter has received reports of underinfusion on the COLLEAGUE Volumetric Infusion Pump. Our investigation has shown that in the presence of an obstruction during IV Administration set loading, certain pump head components can be moved out of alignment, resulting in underinfusion. The amount of underinfusion is typically 5% to 9%, but may be up to 19% below the programmed infusion rate.

To avoid an obstruction during IV administration set loading:

  • Ensure that when loading the IV administration set, the tubing is loaded along the entire length of the tubing channel to avoid a misload or incomplete load situation.
  • Never insert tools or other objects into the tubing channel when attempting to load or unload the administration set.
  • Never use the Manual Tube Release to load or unload the administration set during normal operation.

If you suspect the accuracy of your pump may be affected, perform the Annual Operational Checkout process (which includes an accuracy test) as described in the Service Manual. If the pump fails the accuracy test, remove the pump from service and have the pump inspected by authorized service personnel following the instructions outlined in the COLLEAGUE Volumetric Infusion Pump Service Manual.

Baxter is in the process of verifying changes to the pumping mechanism to better secure these components to reduce the potential for component movement and resulting underinfusion. Baxter will notify you when these changes are available.

Non-Detection of Upstream Occlusion:

As previously communicated in the attached April 18, 2005 Urgent Device Correction Buretrol letter, there is a potential for non-detection of upstream occlusion when Buretrol tubing sets are used with the Colleague Volumetric Infusion pump.

In addition to the condition described above, the following conditions should be avoided as they may result in the pump not detecting an upstream occlusion:

  • Use of a source container which has had all air removed.
  • Incomplete insertion of the spike into the source container.
  • Improper venting of a rigid (glass bottle) or semi-rigid (plastic) container, including Buretrols. If using rigid non-vented containers, refer to the appropriate administration set instructions to determine the correct venting procedure.
  • Unopened air vent above the burette chamber during infusion.

To help ensure upstream occlusions are detected by the pump when using a Buretrol set, do not invert the set and squeeze fluid back into the primary container. Doing this may wet out the vent filter, and obstruct airflow.

Baxter will notify you when the new release of the COLLEAGUE Volumetric Infusion Pump Operator's Manual is available. In the meantime, please immediately disseminate the attachments to all users. If you have any questions about this or any other issues regarding COLLEAGUE pumps, please contact Baxter Medication Delivery Services at 1-800-THE-PUMP (1-800-843-7867).

Please complete the attached reply form confirming your receipt and understanding of this letter and fax it back to Baxter at the number provided on the form. Returning the form promptly will prevent you from receiving a repeat notice. If you provide the COLLEAGUE Volumetric Infusion Pump to other services or facilities, please forward this information, as it is imperative that all end users are notified and confirm receipt of this notification. Additionally, if your pump is serviced by a third party, please forward this information to the service provider.

Health Canada has been notified of this communication.

Baxter Corporation
4 Robert Speck Parkway, Suite 700
Mississauga, Ontario L4Z 3Y4
905.270.1125

Any suspected adverse incident can also be reported to:
Health Products and Food Branch Inspectorate
HEALTH CANADA
Address Locator: 2003D
Ottawa, Ontario K1A 0K9
Tel: The Inspectorate Hotline 1-800-267-9675

The Medical Devices Problem Report Form and Guidelines can be found on the Health Canada web site.

For other inquiries related to this communication, please contact Health Canada at:
Medical Devices Bureau
E-mail: mdb_enquiries@hc-sc.gc.ca
Tel: (613) 957-4786
Fax: (613) 957-7318

Sincerely,

original signed by

Pam Bobbette,
VP Quality,
Baxter Corporation

Attachments: Attachment 1- Battery Usage Guide
Attachment 2 -Infusion Management Guide
Attachment 3 -April 18, 2005 Urgent Device Correction Buretrol Letter

cc: Director of Biomedical Engineering

COLLEAGUE and Baxter are registered trademarks of Baxter International, Inc. 2005FCA35


Attachment 1

Baxter
Colleague
Product Information


Battery Usage Guide


It is imperative that institutions have a contingency plan to mitigate any disruptions of infusions of life sustaining drugs. The interruption or delay of life sustaining therapy may result in potential injury or death. You should consider not using these pumps in situations where a replacement pump is not available or where an interruption in therapy may be life threatening.

Before initially powering on the pump, charge the battery for at least 12 uninterrupted hours. A complete charge may take longer than 12 hours.

The Battery Charge Level Indicator will overstate the remaining battery charge level, and operating time after a "Battery Low" alert can be significantly less than expected, if the pump's batteries are subjected to any of the following conditions:

  • Failure to recharge the batteries for at least 12 uninterrupted hours after occurrence of a "Battery Low" alert or a "Battery Depleted" Alarm.
  • A pump stored with its power cord unplugged, especially with low or depleted batteries.
  • More than one deep battery discharge, as indicated on the pump's "Battery and Pump History" screen.
  • Batteries discharged/recharged more than 68 times as indicated on the pump's "Battery and Pump History" screen.
  • Failure of the "Battery Discharge Test" included in the Colleague Pump Global Service Manual.

Continued pump operation using batteries that have been subjected to any of the above conditions may result in cessation of therapy and/or device failure and loss of configuration memory.

Always do the following to properly maintain the pump batteries:

  • Keep pumps plugged into an AC power outlet at all times except in the event of AC power loss or short-term portable operation.
  • Always store pumps plugged into AC power to maintain the battery charge whenever possible.
  • When a "Battery Depleted" Alarm occurs, immediately plug the pump into a source of AC power. Do not use the pump on battery power again until the batteries have been fully recharged.
  • Charge the battery for a minimum of 12 uninterrupted hours after a "Battery Depleted" Alarm occurs.
  • Never store the pump unplugged and powered on. The batteries may discharge completely, permanently damaging them.
  • Notify Central Supply or other appropriate departments as soon as a pump is removed from patient use so that it can be cleaned and the batteries can be recharged.
  • Pump batteries should be evaluated yearly by Baxter-trained, qualified personnel according to the procedures in the Service Manual, and whenever it is suspected that any of the conditions listed above may have occurred.
  • Baxter-trained, qualified personnel can view the pump's "Battery and Pump History" screen using the instructions provided in the Service Manual.

Attachment 1 to COLLEAGUE Urgent Device Correction Letter Dated 15 Dec 2005


Attachment 2

Baxter
Colleague
Product Information


Infusion Management Guide


It is imperative that institutions have a contingency plan to mitigate any disruptions of infusions of life sustaining drugs. The interruption or delay of life sustaining therapy may result in potential injury or death. You should consider not using these pumps in situations where a replacement pump is not available or where an interruption in therapy may be life threatening.

Generation of Air-Detected Alarm Due to IV Administration Set Tugging

Pulling or tugging on the administration set tubing between the pump channel and the patient may cause a false Air Detected alarm, which will cause the pump to stop infusing. In order to reduce the potential for this situation to occur:

  1. First, select an appropriate length administration set.
  2. Before loading the set into the pump, position the keyed slide clamp at an appropriate location along the tube segment to ensure that there is adequate length of tubing between the patient and the pump to reduce tugging on the set.
  3. Lastly, ensure there is sufficient slack in the tubing between the distal end of the pumping channel and the patient to prevent tube tugging during activities such as moving the patient from one bed to another, or transportation of the patient from one facility location to another.

Conditions That May Affect Pump Delivery Accuracy

Note that flow fluctuations can be caused by unusual conditions or combinations of conditions that may involve, but are not limited to the following:

  • Position of the infusion container
  • Fluid density
  • Positive and negative pressure
  • The environment

Flow fluctuations are most likely to occur when the conditions mentioned above are exacerbated or when the pump is operated in conditions outside of its normal limits.

While the Colleague pump automatically closes the keyed slide clamp, always manually close the regulating clamp on the administration set before removing the set from the pump.

Loading the IV Administration Set

See the following instructions for proper loading of the IV Administration set:

Instructions

  • When loading the IV set, pull the administration set taut and slide it all the way into and along the tubing channel (Figure C and D). The pump pulls in the keyed slide clamp, then loads the administration set into the pumping mechanism. The pump module displays LOADING and then STOPPED.
  • Never insert tools or other objects into the tubing channel when attempting to load or unload the administration set.
  • Never use the Manual Tube Release to load or unload the administration set during normal operation.

If the above instructions are not followed, there is a potential for misaligning the pump chamber components, which may result in an under-delivery condition.

Upstream Occlusion Detection

The pump may not detect an upstream occlusion if one or more of the following conditions exist:

  • Use of a source container which has had all air removed.
  • Incomplete insertion of the spike into the source container.
  • Improper venting of a rigid (glass bottle) or semi-rigid (plastic) container, including Buretrols. If using rigid non-vented containers, refer to the appropriate administration set instructions to determine the correct venting procedure.
  • The air vent above the burette chamber is not open.

To help ensure upstream occlusions are detected by the pump, do not invert the set and squeeze fluid back into the primary container. Doing this may wet out the vent filter, and obstruct airflow.

Fluid Getting into the Pumping Channel

Do not allow fluid to enter the tubing channel. Contact your Baxter Service Center for assistance immediately if fluid enters the tubing channel. The tubing channel should be cleaned as soon as possible by Baxter-trained, qualified personnel to minimize potential difficulties caused by fluid pooling and drying on the mechanism.

Avoidance of the Panel Lockout Button

During storage or patient transport, do not wrap the pump's power cord around the pump tightly enough to accidentally press the PANEL LOCKOUT button on the back. Baxter recommends securing the power cord with the strap provided with the pump instead of wrapping the power cord around the pump.

Verification of Infusion Site and Route

Clinicians are advised to verify the proper route of delivery and that the infusion site is patent.

When using this pump, periodic patient monitoring must be performed to ensure that the infusion is proceeding as expected. The pump is capable of developing positive fluid pressures to overcome widely varying resistances to flow such as resistance imposed by small-gauge catheters, filters, or intra-arterial infusions. Although the pump is designed to stop fluid flow when an alarm occurs, it is neither designed nor intended to detect infiltrations and will not alarm under infiltration conditions.

Piggyback Infusion Feature

When using the piggyback infusion feature ensure:

  • The medication/solution in the secondary source container is compatible with the medication/solution in the primary source container.
  • The secondary administration set is connected to the appropriate injection site on the Continu-Flo administration set.
  • The interruption of the primary infusion is clinically appropriate for the duration of the piggyback infusion.

Attachment 2 to COLLEAGUE Urgent Device Correction Letter Dated 15 Dec 2005


Attachment 3

Baxter

URGENT DEVICE CORRECTION

18 April 2005

RE: INTERLINK BURETROL SOLUTION SET 150ML BURETTE BALL VALVE, Product Code JC7566

Dear Director of Nursing:

Baxter Corporation is providing you with important information concerning the Buretrol set listed above. The BURETROL Set is currently labeled as acceptable for use with the COLLEAGUE Volumetric Infusion Pump. When the set is used with the COLLEAGUE pump, there is a potential for a non-detection of an upstream occlusion under certain conditions. As a result of this information, we do not recommend product codes JC7566 for use with the COLLEAGUE Volumetric Infusion pumps.

If you have any questions concerning the compatibility of other BURETROL Sets with the COLLEAGUE Volumetric Infusion Pump, please contact Baxter Customer Service at 1-888-719-9955.

Please complete the attached reply form confirming your receipt of this letter, and fax it to Baxter using either fax number provided on the form. Returning the form promptly will prevent you from receiving a repeat notice.

If you provide the affected BURETROL Sets to other services or facilities, please forward this communication as appropriate.

Since these products are no longer recommended for COLLEAGUE pump use, you may wish to return the products for credit and order an alternative by calling Baxter Customer Service at 1-888-719-9955.

We apologize for any inconvenience this will cause you and your staff. If you have questions regarding this communication, please contact your Baxter Sales Representative.

Health Canada has been notified of this action.

Sincerely,

original signed by

Pam Bobette
For
Dirk E. Stevens
Vice President, Quality
Medication Delivery Division
Baxter Healthcare Corporation

Buretrol and Baxter are Trademarks of Baxter International, Inc.