A Laser cooled microwave power standardDavid C. Paulusse, Nelson L. Rowell, Alain Michaud We demonstrate the feasibility of a novel microwave power standard based on the electromagnetic interaction with laser-cooled atoms. Under the effect of the radiation, the internal state populations undergo a Rabi flopping oscillation. Measurement of the oscillation frequency allows the determination of the electromagnetic field strength. The measurements are made over a dynamic range of 18 dB and the standard deviation of the measurements is about 0.3 %. The field strength was linear with the Rabi frequency within the accuracy of our measurements. I. IntroductionSINCE the application in the early eighties of concepts related to mechanical effects of light the field now known as laser cooling and trapping has become one of the more studied in experimental physics. Reference [1] gives an index of known links related to quantum optics and laser cooling. It is updated regularly. The first successful applications were in the fields of metrology and spectroscopy, where the atoms in an atomic vapour were probed using laser beams. With these methods the interaction time was increased by many orders of magnitude. Barely a decade after the initial experiments, the absolute measurement of the basic unit of time, the second, has been improved by a factor of five with cooling methods. The new techniques are also expected to improve precision measurements of physical effects such as gravity, magnetic field, rotation of earth, and fundamental constants. This paper summarizes our progress towards using the electromagnetic interaction as a tool in the field of microwave power measurements. To this end we have built a new type of apparatus in which laser cooled Rubidium atoms are subjected to the electromagnetic field. Under the effect of the radiation, the internal state populations undergo a Rabi oscillation. The measurement of the Rabi frequency allows the determination of the electromagnetic field's strength. In the following, we give a short description of the apparatus, we show how it was used to measure the relative power of a microwave signal, and we discuss its resolution and how it could be upgraded to provide an absolute standard for microwave power. II. Atomic Microwave StandardAlthough the standard practical way to measure microwave power is the calorimetric technique, several other approaches have been employed [2]. In one such technique an atomic resonance was used to monitor saturation in the absorption spectrum of a gas (NH3) filling a long section (7 m) of waveguide ([3], [2] section 10.10). When the frequency of the radiation was swept through the resonance (24 GHz), the power could be deduced from the line shape with knowledge of parameters such as temperature and gas pressure (relaxation constant). A. Principle of Operation
1) Preparation of a cold atom sampleFirst, approximately 108 atoms of Rb87 are captured in a standard magneto-optical trap (MOT) [4]. In essence, three Next the magnetic field is ramped to zero, the cooling laser is detuned, and the light intensity reduced. This results in a lower temperature sample which is called "optical molasses" ([4] chap. 7). Subsequently a short pulse of resonant light can be applied to optically pump the atoms in the 5s1/2, F=2 state. Instead of doing this, however, we use a simpler but less efficient expedient, that of simply turning off the repumping light earlier then the cooling laser. 2) Interaction with the Microwave FieldAfter the laser beams has been turned off and the magnetic field nulled, the atoms fall in the interaction zone. At this point all the atoms are in the F=1 state. In this experiment, we apply a resonant pulse of a known width and we measure the population of the level 5s1/2, F=2 at the end of the pulse. We repeat the experiment varying the pulse length each time. It is therefore possible to deduce the value of from by evaluating the period of the relative population of the F = 2 state. In the following analysis, we make the assumptions that there are only two levels, 5s1/2, F=1 and 5s1/2, F=2, their lifetimes are longer than the interaction time, and the electromagnetic field is resonant with the transition. Under the influence of the field, the populations of both levels oscillate as described in the following equations:
Where P1(t) and P2(t) are the relative populations of the states, F=1 and F=2 and the Rabi frequency, b is proportional to the magnetic field amplitude. We also assume that the frequency of the transition is much larger than b, otherwise a more complex model is necessary to describe the mechanism [5]. 3) Detection of Population InversionAfter the atoms have fallen through the interaction zone they are probed using a laser beam. There are many techniques to probe the atoms but the one that we use is the simplest one; i.e. a single resonant laser beam (5s1/2, F=2 to 5p3/2, F'=3 transition) in the trajectory of the atoms. As the atoms fall through this laser beam, their fluorescence light is collected, focused on a photodetector, and the measured intensity is recorded. The amplitude of this time-of-flight signal is proportional to the population of the level 5s1/2, F=2 and the total number of atoms in the cloud. A more Efficient system would use a second laser to probe the atoms in the 5s1/2, F=1 level as well. As this would be a more complicated detection technique to perform, it has not yet been implemented despite the anticipated improvements in signal to noise ratio. B. Description of the Apparatus1) Laser systemThe laser system is based on three laser diodes. The cooling laser is kept in resonance with 5s1/2 F= 2 to 5p3/2 F=3 transition. The output of this laser injects a second laser used as an amplifier. The output power that is sent to the vacuum chamber is approximately 35 mW. The repumping laser, similar to the first one but tuned to the F=1 to the F'=0 is necessary to prevent optical pumping. Its characteristics (power and linewidth) are not critical. 2) Microwave FieldThe geometry of the radiating system is very critical. It has to be simple and symmetrical so that the field distribution can be calculated. Moreover it must have sufficient optical access for the many laser beams. To inject the microwave field we have used an open-ended waveguide (R-70, 6.8345 GHz). This device has been studied [6] and its simple geometry will allow the construction of a mechanically precise device. The microwave signal is generated by a synthesizer (Agilent 83620A) followed by a pulse modulator (Agilent 11720A) and an amplifier (Agilent 8349B). The output power from the amplifier is measured using a thermistor mount, through a 9.5 dB directional coupler and a NBS type IV bridge (Tech USA) and a DC voltmeter (Agilent 3457A). The pulse modulator is turned on for a few seconds and, to compensate for systematic effects such as drift, the power is measured before and after an experiment is run. We should mention here that the measurement of a CW signal is in principle possible with our system. This would, however, require two laser pulses that are spaced by a time interval. The first pulse would perform the optical pumping as described previously. The mechanical effect of the second pulse (F=1 to F'=0) would blow away the atoms that are left in the F=1 state after a certain time. Following this sequence, the total number of atoms as a function of the time interval would show the Rabi fringes. Although this system has yet to be tested, it would allow the measurement of continuous wave signals. 3) Vacuum chamberThe vacuum chamber consists in a glass cube (50 mm x 50 mm x 90 mm) attached to a high vacuum pump and a reservoir filled with rubidium metal. The laser beams, and microwave field are directed onto the atoms through the optical-quality glass plates forming the walls of the cube. To evaluate accuracy as a standard, the dielectric loading of the glass will have to be evaluated or, alternatively, a new system could be built, one in which the waveguide penetrates into the vacuum chamber. For these reasons we have not attempted to evaluate the absolute accuracy of this standard. Instead the experimental results in this paper focus mainly on the resolution and linearity of the system when used as a microwave detector. III. Experimental Results and DiscussionsA. Rabi FrequencyFigure 2 shows four time-evolution plots at different power levels. The height of each dot is proportional to the number of atoms in the F=2 state. The solid curve is derived from a cosine function plus a linear baseline. By virtue of the laser cooling, the interaction time can be as long as 20 ms limited only by the effect of gravity. Also, due to absence of collisions in the cooled atoms, the fringes do not fade significantly, as would be the case for atoms at room temperature.
The noise on those curves is due to the shot-to-shot fluctuation of the total number of atoms in the sample. Such fluctuation, common with MOT's, cannot be reduced significantly. However, the system could be improved by implementing a detection technique in which the readings would be normalized by the total number of atoms in the sample using the two detection laser method (described above). Nevertheless even if the time evolution plots are moderately noisy, the frequency can be estimated with relative accuracy due to the periodic structure of the curves. This shows the potential for this technique. Also we see that any offsets or tilts caused by laser frequency drift or variation in detector performance have little effect on the estimated frequency. Each measurement cycle lasts about 3 seconds resulting in a total measurement time of about 30 minutes for each plot in Fig. 2. The measurement time could be reduced by increasing the background gas density, which depends exponentially on the oven temperature. However this was not found be necessary and the oven was set to room temperature. Also, the complete time-scans are plotted for illustrative purpose only. If the power were known approximately, then the measurement of certain specific points for every cycle would suffice for an accurate frequency estimation. Finally, long-term drift could be measured easily by continuously monitoring the position of the last fringe. B. LinearityWe made several time-scans like the ones shown in Fig. 2, to cover a dynamic range of about 20 dB. The high power end was limited by the maximum power of the source whereas the lowest measurable power was limited by the increasingly long interaction times at low power. In power the range was between 1.5 and 85 mW. A summary of these measurements is plotted in Fig. 3(a). As expected the Rabi frequency is proportional to the field amplitude. Figure 3 (b) shows the relative deviation of the points in Fig. 3(a) from a straight line. It shows that the typical noise on the measurements is about ± 0.3 %. Also we see that, if there is any non-linear dependence, it is smaller than the random noise in the measurements. More precision in the measurement over a wider power range will be necessary to verify the linearity of this device to a higher degree of certainty. Figure 3 - (a: upper) Measured Rabi frequency as a function of the relative field amplitude. The gain of the antenna is not known, but for ease of comparison with figure 2, the horizontal axis units are expressed as a function of the square root of the power sent to the waveguide. (b: lower) Deviation from a straight line in (a) normalised by the value of the highest point. IV. Conclusion and OutlookWe have demonstrated that the precise measurement of the field strength can be done using the electromagnetic interaction with cold atoms. The standard deviation of the measurements was about 0.3 %. The measurements were made over a dynamic range of 18 dB. Under those conditions, Rabi frequency was observed to be linear with microwave field strength. Further investigations and improvements stimulated by the present work include:
References[1] D. Rice, "Quantum Optics and Atom Optics links", Online. Available: http://www.quantumoptics.net. [2] A.Fantom, Radio Frequency and Microwave Power Measurement, London, Peter Peregrinius Ltd., 278 pp., 1990, IEE press, Available: http://www.iee.org [3] S. Hashimoto, and M. Yamaghushi, "Microwave Power Measurement Using Absorption Spectrum of Ammonia Gas," IEEE Trans. Instrum. Meas., vol. IM-25, 4, pp. 348 – 352, Dec. 1976. [4] H. J. Metcalf, P. van der Straten, Laser Cooling and Trapping. New York, Springer-Verlag , 1999, pp. 156 – 162. [5] M. P. Silverman, Probing the Atom. Princeton, NJ: Princeton University Press, 2000, 226 pp. [6] Fred E. Gardiol, "Open-ended waveguides, principles and applications", in Advances in Electronics and Electron Physics, P.E. Hawkes, Ed. New York: Academic Press, 1985, pp. 139-187. Contact: Alain Michaud Low FrequencyP. Filipski, M. Côté This project covers the field of low frequency (below 1 MHz) standards of capacitance, inductance, ac/dc transfer, ac voltage and ac voltage ratio. Also included in this program are dc voltage ratio standards and standards of resistance larger than 1M. Work continues on deriving the NRC unit of capacitance with respect to the ohm (). In the future this will eliminate a dependence on the international community in the determination of the Canadian unit of capacitance, and the impedance scale in general. Preliminary precision measurements of the ac quantised Hall resistance (QHR) have been made. While the ac QHR has been known to exist for some years, the accuracy with which this effect can be used to determine an ac resistance is still unknown.
Work is continuing on the re-determination and extension of our ac voltage scale. During last year NRC took part in an international comparison of ac-dc voltage transfer at lowest level of uncertainty, 3 V at 1 kHz, 20 kHz, 100 kHz, 1 MHz. (CCE comparison 92-3). The results of this comparison are very satisfactory; they showed very good agreement of our highest accuracy standards with those of other leading laboratories. The automation of NRC's two ac/dc voltage transfer systems is nearing completion. The NRC low frequency scale has been extended down to 2 mV, in the 10 Hz - 1 MHz frequency range and up to 1000 V at, 30 kHz. Work is continuing on extending the frequency range of 1000 volt range. The current ac/dc transfer system is also nearing completion. The semi-automated IVD calibration system has been upgraded; we offer IVD calibrations at 1 kHz. Automation and full process control using modern computer techniques has been implemented in the high resistance (1M to 100T) calibration service area. The service allows client's standards to be extensively analyzed as part of the calibration activity and the additional information to be provided as part of the service at no additional costs. Improving calibration services in response to clients' needs by evaluating the development of Hamon ratio devices beyond the range presently available is continuing. Also being researched are bridge technologies based on computer controlled programmable voltage sources. Some of the on-going research has attracted interest from a Canadian high tech company.
R.F. Clark, A.P. Jurkus, D.C. Paulusse This project includes work on RF and microwave standards of voltage, attenuation, impedance, power and noise in the range of 0.1 MHz to 110 GHz. The primary standards of RF power in the frequency range 0.1 MHz to 18 GHz are coaxial calorimeters. The calibration of power meters is carried out by means of a power transfer standard which is first calibrated against the reference calorimeter. The basic standard of waveguide power is a semi-automatic microcalorimeter used to measure the effective efficiencies of waveguide thermistor mounts covering the frequency range 8.2 to 40 GHz. Calibration of working-standard thermal voltage converters and micropotentiometers which are used from 0.1 to 900 MHz is done with the primary power standard calorimeter as the reference. To improve the accuracy, a number of calorimetric thermal converters with calculable rf-dc differences have been built. Accurate calibration of transfer standards of attenuation is available on a fixed frequency basis for coaxial devices over the frequency range 1 MHz to 18 GHz and waveguide devices from 2.6 to 36 GHz. For a reasonably accurate measurement of 'S' parameters obtained at high speed, vector network analyzers are now used which give coaxial frequency coverage from 5 Hz to 50 GHz and waveguide coverage up to 110 GHz. Work on the development of sixport network analyzers included the building of dual sixport systems for the measurement of impedance and attenuation in coaxial lines in the frequency range 0.1 MHz to 18 GHz. Single sixports for power measurement cover the range 0.5 GHz to 18 GHz. The calibration of network analyzers relies on coaxial lines as impedance standards. The inner and outer conductors of these lines are precision machined from solid material, with air only (no supporting beads) being the dielectric. The capacitance per unit length (and characteristic impedance) can be obtained from the diameters of the inner and outer conductor, or more recently, from low-frequency measurement of the capacitance of the line and mechanical measurement of its length. Several lines were measured and the agreement between the two methods was within 0.05%. SET Current StandardJ. Lapointe, B.M. Wood, A.D. Inglis This is a project to study the possible application of Single Electron Tunneling (SET) devices as an electrical current standard. It involves the fabrication of the devices as well as their testing at low temperatures. Single electron tunneling has recently been observed in very small metallic tunnel junctions and in some gated semiconductor devices. This effect only becomes apparent in very small structures (~100 nanometers) and at low temperatures, where the energy e2/2c related to the transfer of a single electron can be the dominant energy scale in the system. Gated circuits with two weak links, also known as electrometers or transistors, are sufficient to observe the charging effect corresponding to a single electron but more complex devices, referred to as turnstiles and single electron pumps, are required to accurately control the transfer of single electrons. In turnstiles and pumps a signal of frequency f produces a current I = ef, making them candidates as an electrical current standard. Attempts are being made to produce SET devices at NRC using both very small aluminum tunnel junctions and silicon-based semiconductor structures, through a collaboration between INMS and IMS. Using facilities of IMS, electrometers were fabricated from metal lines as narrow as 50 nm, using a self-aligning double angle shadow evaporation technique. The devices were measured at temperatures between 4.2 K and 0.3 K in 4He and 3He cryostats using specially designed low-noise electronics. The electrometers clearly showed single electron charging effects, although an apparent gate modulation was observed, presumably caused by background charge fluctuations. The best electrometers had capacitances of about 0.2 fF per junction, with a gate capacitance of about 0.1 fF. Attempts at fabricating silicon-based SET devices have produced no working devices so far. Work continues to improve the yield of our processes and to produce more complex devices. More measurements are planned to understand the origin of the gate fluctuations and to improve on the design of the devices. |
||||||||||||||
![]() |
|