Guidance For Providing Safe Drinking Water in Areas of Federal
Jurisdiction - Version 1
Part 2 - Application of The Federal Framework
4.0 Technical Standards,
Benchmarks and Targets
The following technical standards, benchmarks and targets for
water quality are provided to assist federal departments, water
managers, designers and operators (including those in First Nations
communities) to meet the Guidelines for Canadian Drinking Water
Quality and drinking water-related regulations.
Guidance on how to meet these targets is provided throughout Part
2 of this document. The benchmarks noted below are current as of
July 2005 and may be updated as new information becomes available.
Given the sheer number of federal drinking water supplies and
the wide variety in system size, location, and site-specific concerns,
the guidance in this document is designed to be flexible. In broad
terms, for the purpose of this document, drinking water systems
are defined as large, small, and very small, depending on the size
of the population served (see Section
1.3 or the Glossary for
definitions).
Drinking water sources may be surface water or ground water supplies.
Surface water is defined as water open to the atmosphere and subject
to surface run-off. Groundwater under the direct influence of surface
water (GUDI) is defined as any water beneath the surface of the
ground with (i) significant occurrence of insects or other microorganisms,
algae, organic debris, or large-diameter pathogens such as Giardia
lamblia or Cryptosporidium, or (ii) significant and
relatively rapid shifts in water characteristics such as turbidity,
temperature, conductivity, or pH which closely correlate to climatological
or surface water conditions. Groundwater sources not under the
direct influence of surface water (hereafter referred to as groundwater),
by default, are neither of these two cases.
4.1 Microbiological quality
The most significant parameters to be monitored relate to the
microbiological quality of drinking water. The presence of pathogens
can result in serious health hazards. Indeed, ensuring the microbiological
quality of drinking water is the highest priority for protecting
public health.
For this reason, it is important that federal water quality management
systems address any existing and potential microbiological quality
concerns at the site (e.g., level of treatment, historical
testing results, surface and groundwater vulnerabilities, potential
water quality risks, user activities and local trends), prior to
establishing a routine monitoring regimen.
Microbiological parameters include the indicator organisms E.coli and
total coliforms. System operators may also choose to test for heterotrophic
plate count bacteria in order to better understand the overall
quality of the drinking water. Turbidity and chlorine residuals
should also be monitored in order to ensure the microbiological
integrity of the water, even though these are not specifically
microbiological parameters.
Table 4.1 provides a summary of the recommended sampling frequencies
and locations for the microbiological parameters.
![top of page](/web/20061214093310im_/http://www.hc-sc.gc.ca/images/ewh-semt/arrow_up.gif)
4.1.1 Sampling
frequencies
The recommended sampling frequency for microbiological parameters
depends upon the quality of the source water, the number of water
sources, the past frequency of unsatisfactory samples, the adequacy
of treatment and capacity of the treatment plant, the size and
complexity of the distribution system, the practice of disinfection,
the size of the population served, and sampling history (15).
For federal departments and First Nations communities that collect
and treat their own drinking water, the variables listed above
make it difficult to apply a universal sampling frequency formula.
Table 4.1: Recommended sampling frequencies and locations for microbiological
parameters (See text in Section 4 for detailed guidance based on
system size and unique facilities/situations)
Parameter |
Type of system |
Recommended sampling frequency |
Sampling location |
E.coli
Total Coliforms |
All systems (For guidance on very small systems see 4.1.1) |
In general, 4 samples / month at regular intervals . For
very small systems, case-specific guidance is provided (see
4.1.1) |
Water entering the distribution system and at representative
points in the network |
Existing wells and supplies serving very few individuals,
where the supply has a history of producing water of high bacteriological
quality |
When risk of contamination is highest (See Section 4.1.1) |
After disinfection (where practiced), as water leaves the
pumphouse |
New or renovated wells |
One sample at start-up and then as above |
As water leaves the pumphouse |
Federal buildings receiving municipal water |
1 sample / month (or more/less based on sanitary survey /
vulnerabilities assessment) |
In the building's plumbing |
HPC Bacteria (optional) |
All |
See above guidance |
See above guidance |
Turbidity (See 4.2) |
Surface water & GUDI systems (filtered) |
Continuous or
1 / day to 1 / week |
In the treatment plant at each filter, post filtration |
Surface water & GUDI systems (unfiltered) |
1/day (grab sample or daily average) |
At the source, prior to treatment |
Groundwater systems |
1 / day; or can reduce to 4-12 / year + during high-risk
periods if historical turbidity results consistently less than
1 NTU |
At the source, prior to treatment |
Chlorine residual (See 4.3) |
Disinfected surface water systems |
Continuous or 1 / day (grab sample or daily average) to 1/week |
In the distribution system, after disinfection |
Groundwater systems (if chlorination is used) |
1 / day (See Section 6.3)
|
In the distribution system, after disinfection |
Sampling frequency should be established by the appropriate department
or responsible authority, at the facility or system level, with
proper guidance and after due consideration of local conditions
and sampling history. The following sampling frequencies are offered
as a guide and do not necessarily apply to very small systems.
Generally, for all systems, a minimum of four samples per month
should be taken at regular intervals throughout the month. Sampling
could be scheduled to match provincial guidelines or regulations
where these are more stringent. It is recommended that chlorine
residuals be tested when bacteriological samples are taken.
![top of page](/web/20061214093310im_/http://www.hc-sc.gc.ca/images/ewh-semt/arrow_up.gif)
For very small systems which serve very few individuals,
and where the supply has a history of producing water of high bacteriological
quality, it may be possible to reduce the number of samples analysed (15).
In such cases, samples should be collected and analysed when the
risk of contamination is highest (i.e. there is potential public
health risk) (e.g., spring thaw, heavy rains or dry periods,
and/or when a noticeable deterioration in water quality occurs).
In addition to this, new or renovated wells should be sampled and
analysed at start-up to confirm acceptable bacteriological quality.(18)
Facilities that receive municipal water should
already be receiving microbiologically safe water. Nonetheless,
technical support staff and drinking water monitors should adopt
internal water monitoring procedures and schedules to suit local
conditions.
As a general rule, these facilities should take one bacteriological
sample per month. Actual sampling frequency, number of samples
and range of analyses completed should be based on the sanitary
survey and vulnerabilities assessment. If they demonstrate that
no problems exist and the municipality is sampling at the recommended
frequency (or more), fewer samples could be deemed appropriate.
Conversely, if a greater degree of risk or unknowns exists, the
planned sampling should be adjusted accordingly in order to gather
sufficient data upon which to base corrective actions. If corrective
actions are successful, sampling could ultimately be scaled back
to the recommended frequency according to this document.
If the review of the municipal water quality against federal benchmarks
reveals non-compliance, measures should be taken to protect public
health. Measures could include maintenance (e.g., disinfection
and flushing), additional treatment, or providing an alternative
safe supply of drinking water.
4.1.2 Sampling
locations
Samples should be taken at the point where the water enters the
system (to eliminate the source water as the cause of the adverse
water quality) and from representative points throughout the network,
although not necessarily the same points on each occasion. For
very small systems where there is little or no distribution system,
samples should be taken as the water leaves the pumphouse.
If the water supply is obtained from more than one source, the
location of sampling points in the distribution system should ensure
that water from each source is periodically sampled. The majority
of samples should be taken in potential problem areas: low-pressure
zones, reservoirs, dead ends, areas at the periphery of the system
farthest from the treatment plant and areas with a poor previous
record. (18, 51)
Facilities that receive municipal water should collect samples
at the main or point of entry to the building, in the building's
plumbing system, and in other locations identified in the sanitary
survey and vulnerabilities assessment.
Sampling locations within the building's plumbing system could
be randomly selected at first, to find "hot spots." Over
time, these "hot spots" could become regular sampling
locations. Samples should occasionally be collected from other
locations in the building.
![top of page](/web/20061214093310im_/http://www.hc-sc.gc.ca/images/ewh-semt/arrow_up.gif)
4.1.3 Interpreting
sampling results
4.1.3.1 Total Coliforms and E.
coli bacteria
General:
Microbiological quality is measured by testing samples for total
coliforms and E.coli. Samples should be analysed by an
accredited laboratory or by trained personnel using an accepted
portable laboratory test kit. (See Section
7.3.2.1)
The Guidelines for Canadian Drinking Water Quality state
that the maximum acceptable concentration (MAC) for bacteriological
quality of drinking water systems is no coliforms detectable per
100 mL. Drinking water that fulfills all the following conditions
is considered to conform to this MAC:
- No sample should contain Escherichia coli (E. coli).
The presence of E. coli indicates recent faecal contamination
and the possible presence of enteric pathogens that may adversely
affect human health. If E. coli is detected, the appropriate
agencies should be notified, a boil water advisory should be
issued, and corrective actions taken.
- No consecutive samples from the same sampling location within
a distribution system, or not more than 10% of samples from the
distribution system in a given calendar month, should show the
presence of total coliform bacteria. In systems collecting fewer
than 10 samples per month, no samples should contain total coliforms.
The ability of total coliforms to indicate the presence of faecal
pollution is less reliable than E. coli. However, this
group of bacteria is a good indicator for quality control purposes.
While the presence of total coliforms does not necessarily mean
a boil water advisory must immediately be issued, re-sampling and
corrective actions may need to be taken.
Very small systems with little or no distribution
system:
For very small systems with little or no distribution system,
the following guidance could apply:
- No sample should contain E. coli. As stated above,
the presence of E. coli indicates faecal contamination
and the possible presence of enteric pathogens; therefore the
water is unsafe to drink. If E. coli is detected, a
boil water advisory should be issued and corrective actions taken.
- No sample should contain total coliform bacteria.
The presence of total coliforms does not necessarily mean a boil
water advisory must immediately be issued; however, corrective
actions should be taken.
In non-disinfected well water, the presence of total coliform
bacteria in the absence of E. coli indicates the well
may be prone to surface water infiltration and is therefore at
risk of faecal contamination. In disinfected water systems, the
presence of total coliform bacteria indicates a potential failure
in the disinfection process. In both disinfected and non-disinfected
systems, total coliform detection may also indicate the presence
of biofilm in the well or plumbing system.(18) A
biofilm is a community of micro-organisms attached to a solid surface,
for example the inside wall of a pipe, in an aquatic environment.
Even though the biofilm itself is not a health concern, it could
interfere with analytical testing. Also, it could eventually impede
water flow, potentially leading to deterioration of aesthetic water
quality and ultimately taste and odour problems.
![top of page](/web/20061214093310im_/http://www.hc-sc.gc.ca/images/ewh-semt/arrow_up.gif)
Boil water advisories
For very small systems, boil water advisories should be issued--or
an alternative safe source of drinking water used--in the following
cases:
- If the water supply is not disinfected and is contaminated
with either total coliforms or E.coli.
- If the water supply is disinfected and water leaving the pumphouse
has no chlorine residual or is contaminated with either total
coliforms or E.coli.
It is recommended that a boil water advisory be used only as a
temporary measure while problems are being identified and remediated.
If total coliforms are present with no E. coli present,
it is likely due to the presence of biofilm. If this is the case,
re-sampling and corrective action should be taken. These corrective
actions could include shock chlorination (the addition of a strong
solution of liquid chlorine into a drinking water system to reduce
the presence of microbiological contaminants) and flushing of the
well and/or distribution system.
If there is no evidence to suggest faecal contamination, a boil
water advisory may not be necessary.
A recommended approach to shock chlorination of
wells can be found in Health Canada's "Water Talk: What's
In Your Well? - A Guide To Well Water Treatment And Maintenance",
at http://www.hc-sc.gc.ca/ewh-semt/water-eau/drink-potab/well_water-eau_de_puits_e.html
For more information on issuing and rescinding
boil water advisories, see Section
7.6 and Appendix 7.
4.1.3.2 Heterotrophic Plate Count
Heterotrophic plate count (HPC) bacteria and background colonies,
although not a suitable indicator of the microbiological safety
of water, are a useful indicator of the overall quality of the
water. Testing for HPC bacteria or background colonies is optional.
Effective treatment can reduce concentrations of HPC bacteria
to less than 10 colony forming units per 100mL of water. These
counts can be used for quality control in water treatment plants
and as a measure of quality deterioration in wells, distribution
lines and reservoirs.
A sudden rise of HPC or background colonies in drinking water
collected from a site that has traditionally had low counts should
give rise to concern. If a sample contains greater than usual levels
of HPC or background colonies, the site should be re-sampled and
the disinfectant residual verified. If the repeat sample still
indicates an elevated HPC, the system should be inspected to determine
the cause, and if necessary, remedial action should be taken. (18)
![top of page](/web/20061214093310im_/http://www.hc-sc.gc.ca/images/ewh-semt/arrow_up.gif)
4.2 Turbidity
Federal facilities and facilities in First Nations communities
that treat and supply their own drinking water must monitor their
water for turbidity as it is a strong indicator of water quality.
It is also an important indicator of treatment efficiency and filter
performance in particular.(16) Turbidity
can be measured by on-line turbidity meters, a laboratory, or by
using a test kit.
Turbidity in water is caused by suspended and colloidal matter
such as clay, silt, fine organic and inorganic matter, and plankton
and other microscopic organisms.(16) Control
of turbidity in water supplies is important for both health and
aesthetic reasons. Water that has high levels of turbidity is not
only unappealing to the consumer, but the substances and particles
that cause turbidity can interfere with disinfection and can be
a source of disease-causing organisms.
Turbidity samples must be tested immediately by a trained person
using a turbidity meter that measures in Nephelometric Turbidity
Units (NTU). (48)
The most recent Canadian guideline for turbidity (October 2003)
states that waterworks systems that use a surface water or groundwater
source under the direct influence of surface water (GUDI) should
filter the water to reduce the level of turbidity to as low as
reasonably achievable, with a treated water turbidity target of
less than 0.1 NTU at all times. However, actual levels of turbidity
achieved will vary from technology to technology, so for this reason,
the turbidity guideline is broken down by type of technology.
For guidance regarding situations where the target of 0.1.NTU
is not achievable, and for criteria that describe the circumstances
when filtration is not required, see the supporting document for
the turbidity guideline in the Guidelines for Canadian Drinking
Water Quality.
4.2.1 Guidance
for surface water or GUDI sources that use filtration
For systems that use a surface water or GUDI source and use filtration,
the water that passes through each filter should be continuously
monitored to make sure the filters are operating properly and to
determine when backwashing is necessary. Micro-organisms concentrate
on filters and a breakthrough could result in contaminated water
reaching consumers.
Where continuous monitoring is not feasible, a sample of water
that has not been disinfected should be taken at least once per
day (for large and small systems) or once weekly (for very small
systems) from a point in the system where all filtration has been
completed. Frequency may be increased or decreased depending upon
historical data. A turbidity sample should also be taken at each
filter outlet following filter cleaning or backwashing.
4.2.2 Guidance for
unfiltered systems that use a surface water or GUDI source
For unfiltered systems that use a surface water or GUDI source,
turbidity should be measured at least once per calendar day at
the point in the system directly before the first treatment chemical
is applied. The daily source water turbidity level can be based
on either a single grab sample measurement or the arithmetic average
of all the source water turbidity measurements taken in one calendar
day.(2) Frequency
may be increased or decreased depending upon historical data.
![top of page](/web/20061214093310im_/http://www.hc-sc.gc.ca/images/ewh-semt/arrow_up.gif)
4.2.3 Guidance for secure
groundwater sources
The health-based turbidity guideline does not apply to secure
groundwater sources, i.e., those not under the direct influence
of surface water. Turbidity in these cases is non-organic, should
pose no health threat and should not hinder disinfection (in cases
where disinfection is being done). However, for effective operation
of the distribution system , it is good practice to ensure that
water entering the distribution system has low turbidity levels
of around 1.0 NTU.
Therefore, turbidity sampling once per day is suggested for secure
groundwater supplies. However, if historical data indicates that
turbidity is consistently less than 1 NTU, the frequency of routine
monitoring can be reduced to 4-12 times per year depending on the
number of people served.
That said, turbidity should also be monitored during events such
as spring run-off, heavy rains, and seasonal start-ups.
4.3 Disinfection
targets
Barring system-specific exemptions (as described below), all drinking
water supplies should be disinfected4 to
ensure the safety of the drinking water supply in the treatment
plant. In addition, where disinfection is practised, a residual
of an appropriate disinfectant, typically chlorine or chloramine,
should be present at all times in the distribution system. (See
Section 6.3 for further information on chlorine/chloramine residuals.)
The effectiveness of disinfection can be predicted based on a
knowledge of the residual concentration of disinfectant, temperature,
pH (for chlorine and chloramine), and the time between the moment
the disinfectant is added to the water and the moment the water
arrives to the first customer. This relationship is commonly referred
to as the "contact time" or "CT" concept. CT
is the product of C (the residual concentration of disinfectant,
measured in mg/L) and T (the disinfectant contact time, measured
in minutes). This calculation is used by large drinking water systems
as a tool for ensuring adequate inactivation of organisms during
disinfection (18)
C-T tables can be found within each supporting document for
microbiological guidelines for Canadian drinking water quality,
at http://www.hc-sc.gc.ca/waterquality
Chlorine/chloramine residuals should be tested when bacteriological
samples are taken, as identified above, as well as independently.
Chlorine residuals can be verified by a laboratory or by using
an acceptable test kit (See Section
7.3.2.1).
Additional testing of chlorine residuals could also be done to
routinely monitor the integrity of the distribution system. Note
that residuals are not necessary for in-home plumbing (e.g.,
where point-of-use treatment devices are used).
For more information on chlorine residuals, see Section
6.3.
In the case of a groundwater source not under the direct influence
of surface water, disinfection may not be necessary provided an
annual sanitary survey and vulnerabilities assessment (determined
on a case-by-case basis), is conducted to ensure that the source
is not subject to contamination and that conditions have not changed.
If a comprehensive sanitary survey is conducted following the
methods outlined in Appendix K of the Guidance Manual for Compliance
with the Filtration and Disinfection Requirements for Public Water
Systems using Surface Water Sources (35) then
the survey frequency may be reduced to once every three to five
years, as appropriate. However, sufficient microbiological testing
must still be done to ensure the water is potable.
Note: Even in cases where a sanitary survey or vulnerabilities
assessment suggests that disinfection is not required, periodic
disinfection may become necessary in situations where the microbiological
quality of the water deteriorates. For this reason, disinfection
equipment and supplies, or an equivalent incident response mechanism
(such as an alternative source or boil water advisory), must
be available to deal with potential occurrences.
![top of page](/web/20061214093310im_/http://www.hc-sc.gc.ca/images/ewh-semt/arrow_up.gif)
4.4 Contaminant targets
The Guidelines for Canadian Drinking Water Quality lists
many chemical and physical parameters of concern in Canadian drinking
water supplies. Many of these, though, are only a concern in certain
parts of the country due to site-specific geology or industrial
or agricultural activity. For this reason, it is recommended that
federal departments and, in First Nations communities, water managers
and technical support staff, conduct a baseline chemical analysis
of their water supplies to determine which substances should be
monitored as part of the monitoring program. The baseline chemical
analysis should be done for all drinking water supplies (including
municipal and hauled water).
4.5 Drinking water materials
Drinking water materials are materials that come into contact
with drinking water, from its intake at the source through the
treatment plant and the distribution system and all the way to
the consumer's tap (and beyond).
These materials fall into three general categories: treatment
devices (such as filters and reverse osmosis systems and their
components), treatment additives (such as alum and chlorine) and
system components (such as pipes and faucets). Drinking water quality
concerns from these materials are generally related to:
- Leaching of contaminants from the material into the drinking
water.
- Treatment devices not meeting their manufacturer's claim of
efficiency for removal of specific contaminants.
During the construction, maintenance or replacement of any component
of a drinking water treatment or distribution system, only materials
that meet applicable performance and health-based standards should
be used.
Health Canada does not recommend specific brands of drinking water
treatment devices, but it strongly recommends that consumers look
for a mark or label indicating that the device has been certified
by an accredited certification body as meeting the appropriate
NSF International (NSF) / American National Standards Institute
(ANSI) standard. These standards have been designed to safeguard
drinking water by helping to ensure the material safety and performance
of products that come into contact with drinking water. Certification
organizations provide assurance that a product or service conforms
to applicable standards and must be accredited by the Standards
Council of Canada (SCC). An up-to-date list of accredited certification
organizations can be obtained from the SCC ( http://www.scc.ca).
NSF standards are widely accepted in North America. They reference
and incorporate other relevant standards and protocols as appropriate.
The NSF International web site ( http://www.nsf.org)
has information about both health-based and performance standards
related to drinking water treatment devices including a listing
of certified systems. Standards exist for most drinking water treatment
devices. The two key standards with respect to health effects are
NSF/ANSI Standard 60: Drinking water treatment chemicals-health
effects and NSF/ANSI Standard 61: Drinking water system components-health
effects. NSF/ANSI Standard 61 is often referenced in many of the
standards for drinking water treatment devices to ensure the devices
meet minimum health effects requirements.
Plumbing systems (internal building distribution systems) within
federal buildings and in First Nations communities must be designed
and constructed to meet the National Plumbing Code of Canada. This
may not be possible in situations such as foreign embassies in
other countries. In these situations, minimum sanitary engineering
practices should be met.
Other international standards do exist (e.g., British
Standards International), but do not currently address the health-based
issues related to materials that come into contact with drinking
water and therefore should not be considered equivalent. Any chemicals
and additives used in drinking water treatment processes and/or
the distribution system must meet the applicable health-based standards
established by NSF International.(14)
4 Includes bottled
water
|